Description
An easy-to-use defibrillator designed for professional responders. Large intuitive screen displays graphics and ECG readings and cprMAXTM Technology to minimise CPR interruptions.
Features:
- Large intuitive screen is easy to read
- Easy customisation of settings to meet your protocol needs in the field
- Patented ADAPTIV biphasic waveform provides full range of energy from 150 to 360 joules and is upgradeable
- Energy Selection: Adult 200J > 300J > 360J | Paediatric 50J > 75J > 90J
- Visible and accessible battery gauge to assess battery readiness
- CPR countdown timer provides direction for length of hand-on time based on system protocol
- cprMAXTM Technology minimises CPR interruptions by allowing compressions to continue during AED charging
Specs
All specifications are at 20°C unless otherwise specified.
Waveform: Biphasic truncated exponential with voltage and duration compensation for patient impedance*.
Energy Sequence: User configurable, 150 joules–360 joules. Default energy output settings are 200, 300, 360 joules. 360 joules for every shock thereafter.
Charge Time: With new, nonrechargeable battery pack; 200 joules in less than 7 seconds (360 joules in less than 12 seconds).
3-Wire (Lead II) Monitoring Capability: (If ECG display option purchased). Requires purchase of 3-wire (Lead II) monitoring cable and LIFE-PATCH® electrodes.
Device Software: Field upgradeable. Infant/Child Reduced Energy Defibrillation Electrodes: Reduces selected energy by a factor of 4. Intended for use only with children up to 8 years of age or 25 kg (55 lbs).
Safety Classification: Internally powered equipment IEC 60601-1.
Electrical Protection: Input protected against high voltage defibrillator pulses per IEC 60601-1. *Voltage compensation is limited to the voltage that would result in delivery of 360 joules into 50 ohms.
Device settings
Modes:
- AED – Provides operating capability for basic users.
- Manual – Provides operating capability for advanced users.
- ECG – Provides ECG display capability with 3-wire ECG cable.
- Setup – Allows user to configure the device.
- Data Transfer – Allows user to transfer patient data.
- Auto Test – Provides daily automatic tests of hardware and software. Controls: On/Off, Shock, Menu, Two (2) configurable soft keys. User Defined Options:
- Device ID – Assigns unique identifier to particular device.
- Energy Sequence – User configurable from 150 to 360 joules.
- Flexible Energy – Increases only after a lower energy was unsuccessful.
- Auto Analyze – User can configure device to auto analyze, auto analyze after first shock, or prompt user to push analyze key before each analysis period.
- CPR Time – (Post shock or after no shock advised) – User configurable – 15, 30, 45, 60, 90, 120, 180 seconds.
- Device Date/Time
- Voice Prompt Volume – Allows user to change speaker volume.
- ECG Display – (If option purchased) – Turns display on/off for AED mode.
- Motion Detection – User defined On/Off (default On).
- Service Alert – Audio alarm if the device needs servicing. Configurable on/off.
- Manual Access – (If ECG display option purchased) – Devices configured with an ECG display may be set up to allow user to initiate a charge and shock without analysis.
- cprMAX Technology Settings:
- Initial CPR – User defined time for CPR after first analysis regardless of analysis decision. Can be set to OFF, 15, 30, 45, 60, 90, 120 and 180 seconds.
- Pre-shock CPR – Allows for CPR while device is charging. Can be set to OFF, 15, or 30 seconds.
- Stacked Shocks – (ON/OFF) When Off, allows for provision of CPR after each shock.
- Pulse Check – (Always, After Every NSA, After Second NSA, Never) Allows device to prompt for a pulse check either after each shock, after every NSA pulse check, or never prompt for a pulse check (default Never).
Display
Backlit LCD displays number of shocks delivered, elapsed time, text and graphics of heart rhythm and optional ECG.
Size: 120mm (4.7 in) x 89 mm (3.5 in).
Frequency Response: 0.55 Hz to 21 Hz (-3 dB), nominal
ECG Option:
- Waveform Sweep Speed – 25 mm/sec for ECG, nominal.
- Waveform Viewing Time –Minimum 4 seconds.
- Waveform Amplitude – 1 cm/mV, nominal.
- Heart Rate – 20 to 300 BPM digital display, Display “—” if heart rate is less than 20 bpm. Heart symbol flashes for each QRS detection.
ECG information is received from the adult and Infant/Child electrodes in anteriorlateral or anterior-posterior positions. A 3-wire cable can be used for ECG monitoring (Lead II).
Enviromental
One Hour Operating Temperature (from room temperature to temperature extreme, one hour duration): -20 to 60°C (-4 to +140°F).
Operating Temperature: 0° to 50°C (32° to 122°F).
Storage Temperature: -30° to 60°C (-22° to 144°F) with battery and electrodes (maximum exposure limited to 7 days).
Atmospheric Pressure: 575 hPa to 1060 hPa (4572 to -382 meters; 15,000 to -1253 feet).
Relative Humidity: 5 to 95% (non-condensing).
Dust/Water Resistance: IP55 with battery and REDI-PAK™ electrodes installed (IEC 60529/EN 60529).
Bump: 15 g, 1000 bumps (IEC 600-68-2-29).
Shock: 40 g peak, 15-23 ms, 45 Hz cross over frequency.
Drop: 1 meter drop on each corner, edge and surface (MIL-STD-810F, 516.5, Procedure IV).
Vibration: Random vibration test – MIL-STD-810F, Method 514.5, Category 20; Ground vehicle 3.15 g rms 1 hour per axis.
EMI:
- Radiated – IEC 60601-2-4, IEC60601-1-2, CISPR 11 Class B Group 1.
- Immunity – IEC 60601-2-4, IEC 60601-1-2; IEC 61000-4-2 (Level 4), IEC 61000-4-3, IEC 61000-4-6, IEC 61000-4-8.
Event documentation and communication
Memory Capacity: Dual patient storage. Minimum 40 minutes ECG for current patient. Summarized data for previous patient.
Report Types: Continuous ECG, summary (critical resuscitation events and associated waveforms), event log report (report of time stamped entries reflecting operator and device activity), test log report (self test activity report).
Capacity: Minimum 100 time stamped event log entries.
Data Review: CODE-STAT™ 6.1 Medical Informatics System, DT Express™ 2.1 Information Management System or higher.
Communications: Infrared wireless transfer to personal computer.
Battery and Readiness Display
Note: See operating instructions for information on battery care.
Non-rechargeable Battery:
- Type – Lithium Manganese Dioxide (Li/MnO2), 12.0 V, 4.5 Ah
- Capacity – Typically will provide 440 200-joule discharges or 1030 minutes of operating time with a new battery (370 200-joule shocks or 900 minutes of operating time at 0°C (32°F)).
- Weight – 0.45 kg (1.0 lb)
- Shelf Life – (prior to installation) After the battery is stored for 5 years at 20° to 30°C, the device will provide 48 months of standby life.
- Standby Life – A new battery provides device power for 5 years.
- Low Battery Indicator – At least 30 200-joule shocks or 75 minutes of operating time remain when low battery is first indicated.
Rechargeable Battery:
- Type – Lithium-ion, 11.1 V, 4.8 Ah, 53 Wh
- Capacity – Typically will provide 261 200-joule discharges or 608 minutes of operating time with a new fully-charged battery (247 200-joule shocks or 576 minutes of operating time at 0°C (32°F)).
- Battery Charging Time – Within 4.5 hours
- Weight – 0.45 kg (1.0 lb), maximum
- Standby Life – A new fully-charged battery provides device power for 6 months.
- Low Battery Indicator – At least 30 200-joule shocks or 75 minutes of operating time remain when low battery is first indicated.
Battery Charger:
- Supported Battery – Lithium-ion Rechargeable Battery, 11.1 V, 4.8 Ah, 53 Wh
- Electrical – External Power Supply: 100-240VAC, 50/60Hz
- Temperature – Operating: 0°C to 40°C; Storage: -30°C to 70°C
- Charge Time – Within 4.5 hours
- Charge – Constant Current/Constant Voltage within temperature limits
- Length – 270 mm
- Width – 97 mm
- Height – 92 mm
- Weight – 0.5 kg
Physical Characteristics
Height: 8.7 cm (3.4 in).
Width: 23.4 cm (9.2 in).
Depth: 27.7 cm (10.9 in).
Weight: 3.2 kg (7.1 lbs) with one set of REDI-PAK electrodes and one nonrechargeable battery